Initial Inspection

SAFETY SUMMARY

Initial Inspection

Inspect the shipping container for damage. If there is damage to the container or Cushioning, keep them until you have checked the contents of the shipment for completeness and verified the instrument both mechanically and electrically.

The Performance Tests give procedures for checking the operation of the instrument. If the contents are incomplete, mechanical damage or defect is apparent, or if an instrument does not pass the operator’s checks, notify the nearest Agilent Technologies Sales/Service Office (see page 3).

WARNING

To avoid any hazard, do not perform electrical tests when there are signs of shipping damage to any portion of the outer enclosure (covers, panels, and so on).

WARNING

You MUST return instruments with malfunctions to an Agilent Technologies Service Center for repair and calibration (see page 3).

N3970A UM-1

Line Power Requirements

The Agilent N3970A Power Meter operates at line power when applied to the optional external power supply N3979A 9V DC

Operating Environment

The Agilent N3970A Power Meter can be operated at temperatures between - 10°C and + 55°C and at relative humidity of <95%.

Storage and Shipment

The Agilent N3970A Power Meter can be stored or shipped at temperatures between - 25°C and + 70°C and at relative humidity of less than 95%. Protect the module from temperature extremes that may cause condensation within it.

REQUIREMENT COMPLIANCE

Line Power Requirements N 10149