Agilent Technologies 82A, N5161A Configuring the MXG ATE, Accessing the MXG ATE Web-EnabledPage1

Models: 81A 82A 83A 62A N5161A

1 38
Download 38 pages 49.03 Kb
Page 13
Image 13
Configuring the MXG ATE

Getting Started

Configuring the MXG ATE

Configuring the MXG ATE

NOTE For the N5161A/62A the softkey menus and features mentioned in this guide are only available through the Web- Enabled MXG or through SCPI commands. Refer to Accessing the

MXG (ATE) Web- Enabled Page1, the Programming Guide, and to the SCPI Command Reference.

The MXG ATE N5161A and N5162A, are blank front panel versions of the N5181A and N5182A.

Since the MXG ATE N5161A and N5162A, have no front panel softkeys similar the N5181A and N5182A signal generators, in order to access the features mentioned in this installation guide, two options remain available:

1.SCPI commands (refer to the SCPI Command Reference) or

2.The instrument’s Web- Enabled page1.

Accessing the MXG (ATE) Web-Enabled Page1 2

Access the instrument’s Web- Enabled page by entering its hostname or IP address into the address field of an internet browser. Refer to Figure 2- 1 on page 8 and to the Programming Guide.

NOTE If you have an MXG ATE instrument and you do not know the hostname, you can reset the hostname to the default hostname shipped with the instrument by pressing the LAN Reset hardkey, on the front panel, after the instrument has completed its power up routine.

The following example for accessing the MXG ATE instrument’s Web- Enabled uses the instrument’s predetermined default hostname that the MXG ATE ships with (e.g.

a- <instrument model number>- <last 5 digits of the instrument serial number>).

1.Open an internet browser.

2.Enter the MXG ATE’s hostname:

a- <instrument model number>- <last 5 digits of the instrument serial number>

3.Click Signal Generator Web Control from the menu on the left side of the web page to access the front panel Web- Enabled MXG.

1The SCPI commands and Web- Enabled steps are available for the N5181A and N5182A, too.

2The Web- Enabled MXG ATE web page is titled: “Web- Enabled MXG”, since the MXG ATE is part of the MXG signal generator family.