Agilent Technologies 81A Frequency Range and Accuracy Check, Test Equipment, Recommended Model

Models: 81A 82A 83A 62A N5161A

1 38
Download 38 pages 49.03 Kb
Page 24
Image 24
Frequency Range and Accuracy Check

Operation Verification

Frequency Range and Accuracy Check

Frequency Range and Accuracy Check

The frequency range is tested by determining the frequency accuracy relative to the timebase at the frequency limits of the signal generator. This test can be performed with either a frequency counter or a spectrum analyzer.

Table 3-1 Recommended Equipment for Checking N5161A/62A/81A/82A/83A Frequency Range and Accuracy

Test Equipment

 

Recommended Model

 

 

 

 

 

Frequency Counter

N5161A/62A/81A/82A:

 

 

Agilent 53131A or 53132A with Option 010 and 050 or 124

 

 

 

 

 

Spectrum Analyzer

N5161A/62A/81A/82A

N5183A Option 520

N5183A Option

 

 

 

532/540

 

 

 

 

E4440A

- -

 

 

 

 

E4443A

- -

- -

 

 

 

 

E4445A

- -

- -

 

 

 

 

E4446A

 

 

 

 

E4447A

 

 

 

 

E4448A

 

 

 

 