Agilent Technologies 81A, N5161A, 83A N5162A1/82A Test Procedure, Test Setup, Without Modulation

Models: 81A 82A 83A 62A N5161A

1 38
Download 38 pages 49.03 Kb
Page 29
Image 29
N5162A1/82A Test Procedure

Operation Verification

Checking the Output Power

N5162A1/82A Test Procedure

Test Setup

1.Zero and calibrate the power sensor to the power meter:

2. Connect the equipment as shown below:

3.Preset the signal generator: Press Preset.

4.Turn RF on: Press RF On/Off so that the RF On/Off LED lights.

5.Turn modulation off:

Press Mod On/Off so that the Mod On/Off LED turns off.

Without Modulation

6.Set the signal generator frequency to the first value listed in Table 3- 6:

Press Frequency > 125 > MHz.

1.For the N5161A/62A these softkey menus and features are only available through the

Web- Enabled MXG or through SCPI commands. Refer to the Programming Guide or to the SCPI Command Reference.

7.Set the amplitude to 7 dBm: Press Amplitude > 7 > dBm.

8.Configure the power meter as follows:

a.On the power meter, press the

Frequency Cal Fac button.

b.If applicable, select a power meter channel.

c.Use the arrow keys to enter the frequency at which to measure the power.

9.Measure the output power level.

10.Repeat steps 6 through 9 for the remaining frequencies in the table, and confirm that the power level at each point is within limits.

Table 3-5

N5162A/82A Output Power without

Modulation

Frequency

Amplitude

Limits

(MHz)

(dBm)

(dB)

 

 

 

250

7

±2

 

 

 

338

7

±2

 

 

 

425

7

±2

 

 

 

538

7

±2

 

 

 

675

7

±2

 

 

 

850

7

±2

 

 

 

1075

7

±2

 

 

 

1350

7

±2

 

 

 

1700

7

±2

 

 

 

2150

7

±2

 

 

 

2700

7

±2

 

 

 

3400

7

±2

 

 

 

4300

7

±2

 

 

 

5400

7

±2

 

 

 

23

Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Installation Guide

Page 29
Image 29
Agilent Technologies 81A, N5161A, 83A manual N5162A1/82A Test Procedure, Test Setup, Without Modulation