Checking the Output Power

Operation Verification

Checking the Output Power

Troubleshooting Problems with the Frequency Accuracy Check

Verify the cables are connected correctly.

If you are using a frequency counter, verify that you are using the correct channel for the frequencies you are measuring.

If you are using a spectrum analyzer, verify that the spectrum analyzer is set to external reference.

Checking the Output Power

This test verifies that the CW output power from the signal generator is within defined limits. The following table lists the preferred equipment for this test.

Test Equipment

 

Recommended Model

 

 

Power Meter

Agilent E4418B or E4419A/B E- Series

 

 

Power Sensor,

Agilent E9304A (N5161A/62A/81A/82A/83A)

Input: Type- N (m)

Agilent 8485A

(N5183A Option 520)

 

 

Agilent 8487A

(N5183A Option 532/540)

 

 

 

N5161A/81A Test Procedure on page 22

N5162A/82A Test Procedure on page 23

N5183A Test Procedure on page 25

If this test fails, refer to “Troubleshooting Problems with the Output Power Check” on page 27 for further instructions.

Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Installation Guide