Operation Verification
Running Self Test
Running Self Test1
Self Test is a series of internal tests of signal generator functions. If this test fails, refer to “Self Test Failure” on page 17 for further instructions.
It takes about 5 minutes to run the self test.
Use the following procedure to run self test:
1.Disconnect all external cables, including GPIB, LAN, and USB cables.
2.Preset the signal generator: Press Preset > Utility > Instrument Info > Self Test. The following message appears:
•The current status of the
3.Press Run Complete Self Test.
An activity bar displays on the screen indicating the test progress.
If you press Abort while self- test is running, the following message displays:
•The current status of the
•The current status of the
1.For the N5161A/62A these softkey menus and features are only available through the Web- Enabled
MXG ATE or through SCPI commands. Refer to “Configuring the MXG ATE” on page 7, to the Programming
Guide, and to the SCPI Command Reference.