Agilent Technologies N5980A manual BER and Advanced BER Tab, Gating BER Bar

Models: N5980A

1 44
Download 44 pages 59.92 Kb
Page 28
Image 28

Reference for the User Interface

 

 

 

 

On the advanced page the SFP and the SMA Data pattern can be adjusted

 

NOTE

 

 

independently.

 

 

 

 

 

 

 

SFP Error Insertion Rate:

 

 

 

This drop down list gives a set of Error

Insertion Rates that can be

 

 

inserted for the SFP output.

 

 

 

SMA Error Insertion Rate:

 

 

 

This drop down list gives a set of Error

Insertion Rates that can be

 

 

inserted for the SFP output.

 

 

 

Both the lists contain the same set of Error Insertion Rates. The following

 

NOTE

 

 

screen shows this list.

 

 

 

 

 

Trigger Out:

The Trigger Out lists a set of clock triggers. The list is shown below:

Error Detector

Gating Time:

The Gating Time can be set in terms of hour, minute, and seconds.

The maximum gating time is 168 hours 59 minutes and 59 seconds.

BER and Advanced BER Tab

BER

Gating BER Bar

This BER Bar indicates the Bit Error during the Gating Period. Once the Gating Period is complete this Bar turns green if the test passes (Gating BER is 0.0), and turns red if the test fails (Gating BER > 0.0).

28

N5980A User Guide

Page 28
Image 28
Agilent Technologies N5980A manual BER and Advanced BER Tab, Gating BER Bar