Programming Reference

Trigger Output

The trigger output is accessible via the root node :SOUR3

Pattern

:SOUR3:PATT

Syntax :SOUR3:PATT?

:SOUR3:PATT CLK2CLK4CLK8CLK10CLK16CLK20 Description Either sets or gets the pattern of the trigger output.

Available Patterns

Name

Parameter

Clock / 2

CLK2

Clock / 4

CLK4

Clock / 8

CLK8

Clock / 10

CLK10

Clock / 16

CLK16

Clock / 20

CLK20

Error Detector

The Error Detector channel is accessed via the root node :SENS1

Detected Pattern

:SENS1:PATT?

Syntax :SENS1:PATT?

Description Gets the pattern that is currently used on the Error Detector to compare it with the incoming data stream. This might change when the Error Detector detects a very high error ratio, and tries to synchronize with the incoming data.

Response

Error Detector Pattern

 

Response

PRBS 2^7-1

PRBS7

PRBS 2^15-1

PRBS15

PRBS 2^23-1

PRBS23

PRBS 2^31-1

PRBS31

Error Detector has lost pattern

 

UNKNOWN

synchronization.

 

 

38

N5980A User Guide

Page 38
Image 38
Agilent Technologies N5980A manual Trigger Output, Detected Pattern, SOUR3PATT, SENS1PATT?