Programming Reference

Synchronization mode

Syntax

Description

Available

Synchronization

Types

:SENS1:SYNC:TYPE

:SENS1:SYNC:TYPE

AUTOSFPSMAPRBS7PRBS15PEBS23PRBS31

:SENS1:SYNC:TYPE?

Either sets or gets the synchronization type. The synchronization type defines which patterns will be used to synchronize when a synchronization loss is detected by the Error Detector.

Parameter

Synchronization behavior

 

 

AUTO

The Error Detector will use all available

 

patterns to synchronize with the incoming

 

data stream.

SFP

The Error Detector will use only the

 

pattern that is set for the SFP output.

SMA

The Error Detector will only use the

 

pattern that is set for the SMA output.

PRBS7

The Error Detector will only use a PRBS

 

2^7-1 data stream.

PRBS15

The Error Detector will only use a PRBS

 

2^15-1 data stream.

PRBS23

The Error Detector will only use a PRBS

 

2^23-1 data stream.

PRBS31

The Error Detector will only use a PRBS

 

2^31-1 data stream.

Actual Bit Error Rate

:SENS1:FETC:ERAT:DELT

Syntax :SENS1:FETC:ERAT:DELT?

Description Gets the bit error ratio of the last update interval (approx 200ms).

Actual Number of Errors

:SENS1:FETC:ECO:DELT

Syntax :SENS1:FETC:ECO:DELT?

Description Gets the number of errors that have been counted during the last update interval (approx. 200ms).

N5980A User Guide

39

Page 39
Image 39
Agilent Technologies N5980A manual Synchronization mode, Actual Bit Error Rate, Actual Number of Errors