Reference for the User Interface

BER Bar

The BER bar displays the BER calculated during a period of 200 ms.

Error

This indicator turns red when errors are detected.

ƒStable errors are caused by the error add function: it turns SFP/SMA Error Insertion Rate off.

ƒVariable and high errors may be caused by faulty connectors/cables. BERT connected to your device shows:

ƒStable or variable errors caused by your device. Correct the problems with your device.

ƒIf the problems are not clear, then accumulate measurement data and analyze the results.

Sync Loss

This indicator turns red when the measured BER is higher than the sync threshold of 0.1.

ƒThe ED pattern does not match the incoming data pattern.

ƒPatterns have lost sync.

ƒThe device connected to the BERT has inverted the data pattern.

ƒThe device has changed the data pattern.

 

Error Detector can sync only PRBS patterns. Clock patterns cannot be measured

NOTE

 

by the error detector, and will result in loss of synchronization.

 

 