Agilent Technologies Agilent 6031A 1. Performance Specifications, Agilent Technologies Model

Models: Serials 2934A-01825 to 01829 3023A-01925 Serials 2934A-01821 to 01825 3026A-01921 and above Agilent 6032A

1 129
Download 129 pages 812 b
Page 16
Image 16
Table 1-1. Performance Specifications

Table 1-1. Performance Specifications

 

 

Agilent Technologies Model

 

6033A

DC Output: Voltage, current and power spans indicate range

Volts

0-20 V

 

over output may be varied using front panel controls.

Amps

0-30 A

 

 

Maximum Power

200-240 W

Load Effect (Load Regulation) Voltage load effect is given for a load

Voltage

0.01%

+ 2 mV

current change equal to the current rating of the supply. Current load

 

 

 

effect is given for a load voltage change equal to the voltage rating of the

Current

0.01%

+ 9 mA

supply.

 

 

 

6038A

0-60 V

0-10 A

200-240 W

0.01% + 3 mV

0.01% + 5 mA

Source Effect (Line Regulation): Given for a change within the rated line voltage for any output within the rated output voltage, current and power of the supply

PARD (Ripple and Noise): Measured at any line voltage and under any load condition within rating (rms 10 Hz to 10 Mhz/p-p 10 Hz to

20 MHz)

Load Effect Transient Recovery: Maximum time required for output voltage to recover with the specified band around the nominal output voltage following a 10% step change in output current while operating in the constant voltage mode

Programming: (25 ± 5°C) Given for control of the output over the GP-IB or with front panel controls

Remote Voltmeter: (25 ± 5°C) Refers to data read back to the controller over the GP-IB

Remote Ammeter: (25 ± 5°C) Refers to data read back to the controller over the GP-IB.

Front Panel Voltmeter: (25 ± 5°C)

Front Panel Ammeter: (25 ± 5°C)

Voltage

0.01%

+ 1 mV

 

 

 

Current

0.01%

+ 6 mA

 

 

Voltage

3 mV/30 mV

 

 

Current

30 mA/1

Time

1 ms

 

 

 

 

Level

50 mV

 

 

 

Voltage Accuracy

0.035% + 9 mV

Current Accuracy

0.15%

+20 mA

Accuracy

0.07%

+ 6 mV

 

 

Accuracy

0.3% + 25 mA

 

 

Range

2 V, 20 V, 200 V

Accuracy

0.07% +6 mV, 11

 

mV, 56 mV

Range

20 A, 200 A

Accuracy

0.3% + (20 mA, 65

 

mA)

 

0.01% + 2 mV

0.01% + 2 mA

3 mV/30 mV

5mA/1

1ms

75 mV

0.035% + 40 mV

0.085% + 10 mA

0.07% + 50 mV

0.2% + 11 mA

2 V, 20 V, 200 V

0.07% + 50 mV, 55 mV, 100 mV)

20 A

0.2% + 11 mA

Table 1-2. Supplemental Characteristics

 

Agilent Technologies Model

 

 

6033A

Programming: (25 ± 5°C) Given for control of the

 

Voltage Resolution

5 mV

output over the GP-IB or with front panel controls

 

Current Resolution

7.5 mA

Remote Voltmeter: (25 ± 5°C) Refers to data read back to the controller

Resolution

5 mV

over the GP-IB

 

 

 

Remote Ammeter: (25 ± 5°C) Refers to data read back to the controller

Resolution

7.5 mA

over the GP-IB.

 

 

 

Front Panel Voltmeter: (25 ± 5°C)

 

Resolution

5 mV, 10 mV, 100

 

 

 

mV

Front Panel Ammeter: (25 ± 5°C)

 

Resolution

10 mA, 100 mA

Maximum AC Input Current: +6% -13%(48-63) Hz

 

100 Vac (Opt.100)

6.0 A

 

 

120 Vac (Std.)

6.5 A

 

 

220 Vac (Opt.220)

3.8 A

 

 

240 Vac (Opt.240)

3.6 A

Temperature Coefficient: Output change per degree Celsius change

Voltage

50 ppm + 0.6 mV

in ambient following 30 minute warm-up.

 

Current

100 ppm + 2 mA

Drift (Stability): Change in output (dc to 20 Hz) over 8-hour internal

 

Voltage

0.02 % + 1 mV

under constant line, load, and ambient following 30-minute warm-up

 

Current

0.03% + 10 mA

Programming Response Time: The maximum time required

 

Settling Band

5 mV

to change from zero volts to full scale voltage or from full

Up

Full Load

100 ms

scale voltage to 2 volts ( 5 volts for Agilent 6035A) and settle

 

No Load

100 ms

within the specified band. Full load is defined as the

Down

Full Load

200 ms

resistance equal to Vp1/Ip1. Light load is as specified

 

Light Load

500 ms (50 Ω)

Overvoltage Protection: Trip voltage adjustable via front

 

Range

0-23 V

panel control using the Display OVP function

 

Resolution

100 mV

 

 

Accuracy

0.3% + 200 mV

Typical input power at rated output power: (see point P2 on Figure 1-1)

 

340 W

6038A

15mV

2.5mA

15mV

2.5 mA

15 mV, 15 mV, 100 mV

10mA

6.0A

6.5A

3.8A

3.6A

50ppm + 3 mV

90ppm + 0.3 mA 0.02% + 2 mV 0.03% + 3 mA

15mV

150ms

120ms

150ms

750ms (400 Ω) 0-63 V

100mV

0.25% + 300 mV

325W

16General Information

Page 16
Image 16
Agilent Technologies Agilent 6031A, Agilent 6030A 1. Performance Specifications, Agilent Technologies Model, 6033A, 6038A