Calibration

4.Use the and buttons to enter an Applied value that exactly equals the concen- tration stamped on the CO cylinder.

The calibration range is from 250 to

11,000 ppm. An attempt to calibrate out- side this range will cause the message “Bad Calibration Wrong CAL Entry” to appear in the following step.

CALIBRATE CO-HI

Measured

3895 PPM

Applied

4000 PPM

06/28/06 09:45:30 AM

PRINT

5. Adjust regulator for a flowmeter indica-

F1

F2

F3

tion of approximately 2 SCFH. Wait until

 

 

 

the Measured reading stabilizes (approximately 3 minutes), and then press ENT to calibrate the CO Measured value to that of the Applied value­ . The message “Good Calibration” should briefly appear followed by the CALIBRATION LIST screen being re-displayed.

If the sensor’s output is low, but still usable, then the message “Good Calibration WARNING Low Sensor” will appear. The sensor will now be marked as being Low in the DIAGNOSTICS screen.

If the sensor’s output is too low to be usable, then the message “Bad Calibration Sensor End of Life, Entry Not Saved” will appear followed by the CALIBRATION LIST screen being re-displayed. The sensor will now be marked as being BAD in the DIAGNOSTICS screen.

6. Turn OFF regulator and remove gas cylinder.

Leave Top

Port Open

Parts:

1.Gas Cylinder

2.Regulator *

3.3/16" I.D. Tubing *

4.Tee *

5.Flowmeter *

6.Barbed Hose Connector *

* Part of Calibration Kit 24-7059

GAS Port

POWER T-STACK T-AIR

P GAS

Figure 5-2. Gas Sensor Calibration Equipment

5-14

Instruction 24-9448

Page 80
Image 80
Bacharach 24-9448 manual Gas Sensor Calibration Equipment

24-9448 specifications

The Bacharach 24-9448 is a sophisticated instrument designed for detecting and measuring hazardous gases, particularly in environmental and industrial settings. Renowned for its accuracy and reliability, this portable gas analyzer serves a crucial role in ensuring safety and compliance in various applications, including HVAC, refrigeration, and environmental monitoring.

One of the standout features of the 24-9448 is its advanced sensor technology. The device utilizes electrochemical sensors that provide real-time readings of specific gas concentrations, including carbon monoxide, carbon dioxide, and other toxic gases. These sensors are designed to be sensitive and selective, ensuring they deliver precise measurements without interference from other substances in the environment. This is particularly important in industrial settings where multiple gases may be present.

The Bacharach 24-9448 also boasts a user-friendly interface with a clear display that makes it easy for operators to interpret results quickly. The unit features a simple menu navigation system, allowing users to access different functions without hassle. This design consideration enhances usability, especially in high-pressure situations where every second counts.

Portability is another key characteristic of the 24-9448. Weighing in at just a few pounds, it is easy to transport and operate in various locations, whether it be on a job site or in an environmental monitoring scenario. The robust housing is designed to withstand the rigors of fieldwork, ensuring longevity and durability despite challenging conditions.

Moreover, the device includes data logging capabilities, enabling users to store and later review measurements. This feature is essential for compliance documentation and for analyzing trends over time. Users can download data for further analysis, aiding in thorough reporting and decision-making processes.

In addition to its core capabilities, the Bacharach 24-9448 offers flexibility with adjustable calibration settings, accommodating different gas types and concentration ranges. This adaptability ensures it can meet the specific needs of diverse applications, making it a versatile tool in the field.

Overall, the Bacharach 24-9448 is an essential instrument for professionals in various sectors. Its combination of advanced sensor technology, user-friendly design, portability, and data logging capabilities solidifies its position as a leading choice for gas detection and measurement. Whether for routine checks or emergency responses, this analyzer provides the reliability that ensures safety in potentially hazardous environments.