Troubleshooting
7.3 Error Messages Displayed After Warm-Up
If there were problems detected during
If problems were detected, the analyzer will not automatically switch to the Combustion Test HOLD screen after
The following is a list of the error messages that may appear following
Low Battery – Battery voltage is low. Only several minutes of operating time remain. Replace or charge batteries per Section 3.2.
O2 No Sensor – Oxygen sensor not installed, or has become unplugged from its electrical connector. Install oxygen sensor per Figure
O2 BAD Sensor – The oxygen sensor’s output is below 24 mV, signifying that the sensor is depleted and needs replaced. Refer to Section 6.3.
No Smart Sensors – No gas sensors are installed. Install sensors per Section 6.3.
Pressure Overrange – Pressure sensor measured a pressure that was outside the range of ±3 inwc during
Ts Error – The probe’s thermocouple is not connected to the analyzer’s T- STACK connector.
Ts Overrange – The temperature measured during
Ta Overrange – The temperature measured during
UNCALIBRATED SENSORS – List of sensors that do not have a calibration history stored on their printed circuit board. Calibrate each of the sensors listed per Section 5.