8.Verify that the unit is not in a purge cycle. This can be done by verifying a flow of air into zone 1. Read the (PRES) reading in the diagnostic screen (write down reading).
9.Connect the tubing from Reference Gas regulator to Zone 1 on the manifold of HGM 300. Note: The Fault light on HGM300 may illuminate until gas starts to flow from the reference bottle to unit. The pressure should be below the original reading (if reading does not drop the unit may be in the purge cycle. If so wait until purge cycle ends then continue). Adjust the variable regulator slowly clockwise until pressure is the same as original reading. Note: You may have to block or partially block the relief valve on the variable pressure regulator in order to achieve the appropriate pressure. Allow about 1 minute to settle then read the first PPM value in the diagnostics screen. The reading should be within
10.Close reference by adjusting variable regulator counter clockwise until fully closed. Reset zone 1 information (distance, refrigerant, zone hold time) as desired for normal operation.
11.If reading in step 9 is not within the
- Verify pressure from step 8
If the unit still does not measure within
Rev 3/2/09 | 17 |