ADDITIONAL NOTES ON “NO FLOW” PROBLEMS (THE FLOW RATE OF THE PUMP IS 2L/MIN)

A DIFFERENCE OF LESS THEN 0.1 PSIA BETWEEN AMBIENT AND PRESSURE WILL CAUSE A “N0 FLOW” FAULT.

A VACUUM OF 0.7 PSIA OR LESS BELOW PRESSURE WILL CAUSE A “N0 FLOW” FAULT.

TO OBSERVE PRESSURE READINGS ON THE DIAGNOSTIC SCREEN ZONE BY ZONE, YOU MUST PUT A ZONE INTO”ZONE HOLD” THEN GO TO THE DIAGNOSTIC SCREEN AND VERFY THAT YOU ARE OBSERVING THE PROPER ZONE AT THE TOP OF THE SCREEN.

IN LINE FILTER

PUMP

“T” CONNECTOR

PRESSURE SENSOR

I/R BENCH

SOLENOID

MANIFOLD

STEP”A”

PINCH TUBING

STEP “B” PINCH

TUBING HERE

EXHAUST

WATER TRAP

CHECK THAT “FLOAT” IS FREE

ENSURE THAT DRAIN PETCOCK IS TIGHT (TURN CLOCKWISE TO TIGHTEN

Rev 3/2/09