14. Installation
Image14-18
14.10Automatic startupWhat can be done ?
Thea utomaticst artup allowsimmediate restart of the projector after a power failure (breakdown), i.e. withoutpassing through the
standby state.
Theprojec tor restarts at power resume and recovers the previous settings (previous source,...).
Thisfunction can be disabled if undesired or inadequate for safety reasons, etc.
CAUTION:If the Automatic startup function is enabled one must be aware of the fact that it involves safety
precautions
Makesure that the projector (or the operators!) willnot be affected by altered environmental conditions when
restarting at power resume.
How to enable/disable the Automatic startup?
1. Press MENU to activate the Toolbar
2. Press →to select the Installation item
3. Press ↓to Pull down the Installation menu
4. Use ↑or ↓to select Automatic startup
5. Press →to pull down the menu
6. Use ↓or ↑to enable/disable the automatic startup (image 14-19)
7. Press ENTER
120 R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004