10. Settings
Image10-1
Thecustom color tem perature allows the selection of a user defined color temperature set in the Adjustment
menu.
10.1.3 Color temperature in a closed loop medical visualizationsystem
Color temperature in a closed loop medicalvisualization system
If, however the projector is part of a typical closed loop medical visualization system (Barco Dicom®Theatre ), color temperature
willbe set by the control & calibration system (MediCal ®Pro) along with a display function forming a pre-set value.
Thepreset (formed by color temperature and display function) values are checked on a regular basis to verify if the projected image
iswithin a tolerance with respect to consistency, in the case it isn’t, a calibration process will be launched.
How to select a color temperature ?
1. Refer to the system manual
OSD interaction
WhenapresetisselectedonMediCal
®Pro, the selected color temperature (as well as the display function) will be marked on the
projector’s OSD menu.
Full white
Clearbase
Bluebase
Video(6500K)
Computer(9300K)
Custom
Image10-2
Thecompletion of the calibration procedure will in the same way be signaled by a separate logo (as well as for the display function).
68 R5976741 MGP15 MEDICAL GRADE PROJECTO R 28/04/2004