Important Important Things to Know

Important Things to Know:

Your CONTOUR TS Blood Glucose Meter is specific for glucose and has been referenced to plasma/serum values. Methods calibrated to plasma/serum can be 9 to 15% high- er than the results of whole blood glucose assay methods.

Always keep the test strips in the original bottle. Firmly snap the lid closed immediately after removing a test strip. The bottle is especially designed to provide a dry envi- ronment for the test strips. Extended exposure to room humid- ity from leaving the bottle open or not storing the test strips in the original bottle will damage your test strips. Do not expose test strips to direct sunlight.

Avoid excessive “milking” of your finger when forming a drop of blood and test immediately after forming the blood drop.

DO NOT drop blood directly on the flat surface of the test strip. Your test strip is designed to easily draw the blood in through the sample tip.

DO NOT press the test strip against your finger. This may block the sample tip.

DO NOT use a test strip that appears damaged or has been used. Test strips are for single use only.

IMPORTANT NOTE: Your meter has been preset to display results in mg/dL (milligrams of glucose per deciliter). Results

in mg/dL will never have a decimal point (e.g., ); results in mmol/L (millimoles of glucose per liter) will always have

a decimal point (e.g., ). Check your display to ensure that results are being displayed in the expected units; if they are not, contact your authorized Bayer Diabetes Care representative.

6

Need Help? Call toll free: 1-800-348-8100

Page 8
Image 8
Bayer HealthCare TS manual Important Important Things to Know