Trouble-shooting

Error message

Situation

Solution

 

 

 

 

No probe cover is attached.

Attach new, clean probe cover.

 

 

 

 

 

 

 

(POS = position error)

The infrared monitor cannot find a temperature equilibrium and allows no measurement.

Clear the display by pushing the «Start» button once.

Ensure that the positioning of the probe is correct and remains stable.

(flashing)

Ambient temperature is not within the allowed operating range

50–104 °F (10–40 °C) or changing too rapidly.

Allow the thermometer to remain for 30 minutes in a room where the temperature is between 50 and 104 °F / 10 and 40 °C.

Temperature taken is not within typical human temperature range (93.2–108 °F (34–42.2 °C).

HI = too high

LO = too low

Make sure the probe tip and lens are clean and a new, clean probe cover is attached. Make sure the thermometer is properly inserted into the ear canal.

Then, take a new temperature.

 

 

System error

Wait 1 minute until the thermo-

 

 

 

meter turns off automatically,

 

 

 

 

 

 

then turn on again.

 

 

 

 

 

 

 

 

If error persists,

... reset the thermometer by re-

 

 

 

moving the batteries and putting

 

 

 

 

 

 

them back in.

 

 

If error still persists,

... contact Welch Allyn at

 

 

 

1-800-535-6663

 

 

 

(in Canada at 1-800-561-8797).

 

 

 

 