Cleaning and disinfecting the Stimuplex® HNS 11

Use only soft, damp cloths to clean and disinfect the Stimuplex® HNS 11 and the electrode cable. Water, soap or white spirits are especially suitable for this. Ensure that no moisture penetrates the Stimuplex® HNS 11.

Use only wet disinfection, no spray disinfection! Avoid condensation!

White spirit or commercially available methanol-free disinfectant in an ethyl alcohol base can be used for disinfection.

Caution!

The following substances must not be used:

trichloroethylene, acetone, methyl ethyl ketone, benzene, metha- nol, cellulose thinner, 1-propanol and 2-propanol.

Maintenance and safety checks

Check the serviceable condition of the Stimuplex® HNS 11 and ac- cessories on each occasion before use. A defective instrument must not be used.

Electromedical equipment may only be repaired by the manufac- turer or by an organization expressly authorized by the manufac- turer. An order for repairs must be accompanied by a detailed description of the fault (acc. separate Technical Service Manual).