26

Appendix D

Bare and insulated needles

Stimulation current

[mA]

Distance from the nerve [mm]

Stimuplex® D and Contiplex® D needles (DP: DE 3919666)

Current density

Bare,

uncoated

needles

Coated needles

bevel uncoated

Stimuplex® A Contiplex® A

Stainless steel needle

Plastic insulation

Only 0.2 or 0.05 mA

 

stimulation current required

blunt

 

Stimulation

30°

 

current

Fully insulated

[mA]

needle with

 

bare tip

 

Stimuplex® D

 

Contiplex® D

Distance from the nerve [mm]

Threshold currents down to minimum 0.05 mA (with 1.0 ms pulse width) and minimum 0.2 mA (with 0.1 ms pulse width) allow reliable location of nerves.

pointed 15°

Page 26
Image 26
Braun HNS 11 user manual Appendix D, Only 0.2 or 0.05 mA Stimulation current required