The LED displays | 9 |
|
|
All LEDs directly assigned to the buttons are used to monitor the selected operating mode. Since LEDs require relatively high power, the Stimuplex® HNS 11 switches them off after 8 seconds (see p. 6 Section 7).
The individual displays have the following significance:
Current display mode
The LCD display shows the stimulation current set ( ).
The LCD display shows the stimulation current actually flowing (). In addition, a ”+” appears in front of the current displayed.
Current adjustment range
Setting range from 0.00 to 5.00 mA:
Use the controller to dial a stimulation current up to 5 mA. The current is displayed in 0.02 mA digits.
Setting range from 0.00 to 1.00 mA:
Use the controller to dial a stimulation current up to 5 mA. The current is displayed in 0.01 mA digits.
Pulse width
Pulse width 1.0 ms: duration of stimulation pulse is 1.0 ms.
Pulse width 0.3 ms: duration of stimulation pulse is 0.3 ms.
Pulse width 0.1 ms: duration of stimulation pulse is 0.1 ms.
Pulse frequency
Pulse frequency 2 Hz (two pulses per second)
Pulse frequency 1 Hz (one pulse per second)