The LED displays

9

 

 

All LEDs directly assigned to the buttons are used to monitor the selected operating mode. Since LEDs require relatively high power, the Stimuplex® HNS 11 switches them off after 8 seconds (see p. 6 Section 7).

The individual displays have the following significance:

Current display mode

The LCD display shows the stimulation current set ( ).

The LCD display shows the stimulation current actually flowing (). In addition, a ”+” appears in front of the current displayed.

Current adjustment range

Setting range from 0.00 to 5.00 mA:

Use the controller to dial a stimulation current up to 5 mA. The current is displayed in 0.02 mA digits.

Setting range from 0.00 to 1.00 mA:

Use the controller to dial a stimulation current up to 5 mA. The current is displayed in 0.01 mA digits.

Pulse width

Pulse width 1.0 ms: duration of stimulation pulse is 1.0 ms.

Pulse width 0.3 ms: duration of stimulation pulse is 0.3 ms.

Pulse width 0.1 ms: duration of stimulation pulse is 0.1 ms.

Pulse frequency

Pulse frequency 2 Hz (two pulses per second)

Pulse frequency 1 Hz (one pulse per second)