TROUBLESHOOTING (cont.) |
|
|
PROBLEM | PROBABLE CAUSE | REMEDY |
Dripping from sprayhead | 1. Syphon system | The brewer must be level or slightly |
|
| lower in front to syphon properly. |
| 2. Lime | Inspect the tank assembly for ex- |
| CAUTION - Tank and tank compo- | cessive lime deposits. Delime as |
| nents should be delimed regularly | required. |
| depending on local water condi- |
|
| tions. Excessive mineral |
|
| on stainless steel surfaces can ini- |
|
| tiate corrosive reactions resulting |
|
| in serious leaks. |
|
3.Solenoid valve (ST, STF, STA & STFA)
Remove the solenoid valve and clear any obstructions. Rebuild or replace the valve if necessary.
See page 18
Water flows into tank continuously | 1. | Timer (ST, STF, STA & STFA) | Refer to Service - Timer for testing |
(ON/OFF Switch "ON") |
|
| procedures. See page 22 or 24 |
| 2. | Start switch | Refer to Service - Start Switch for |
|
|
| testing procedures. See page 19 |
| 3. | Liquid level board | Refer to Service - LiquidLevel |
|
|
| Board for testing procedures. See |
|
|
| page 26 |
| 4. Relay |
Water flows into tank continuously | 1. Solenoid valve (ST, STF, STA & |
(ON/OFF Switch "OFF") | STFA) |
Refer to Service - Relay for testing procedures. See page 28
Remove the Solenoid Valve and clean any obstruction. Rebuild or replace the valve if necessary. See page 18
Beverage overflows dispenser | 1. | Dispenser | The dispenser must be completely |
|
|
| empty before starting a brew cycle. |
| 2. | Timer (ST, STF, STA & STFA) | Refer to Service - Timer for testing |
|
|
| procedures. See page 22 or 24 |
Page 10 | 29251 052500 |
|