4 Basic Operations
Alignment
1Align the retinal camera with the eye to be photographed.
CAUTION | When adjusting the position of the retinal camera in the |
| unit closer to the patient while looking at the patient’s eye from the side. The patient’s eye may be |
| injured if the objective lens makes contact with it. |
|
|
1.Move the stage until the eye to be photographed appears in the observation monitor.
2.Adjust the center of the patient's pupil and the pupil alignment circles.
3.Have the patient stare at the green blinking internal eye fixation lamp.
Pupil alignment circles
During examination of the anterior segment of a patient who is extremely near- or
• To move the main unit slightly to the left or right, tilt the operation lever to the left or right.
• To move the main unit slightly to the front or back, tilt the operation lever to the front or back.
• To move the main unit up or down, turn the main unit vertical movement ring.
When the vertical position cannot be fully adjusted by using the operation lever
Adjust the height of the chin rest with the CHIN REST button to raise or lower the patient's face.