7Troubleshooting

If any of the symptoms listed here occurs or if a warning appears in the Rics NM 2 while using the CR-2, try the relevant remedy described below.

If these remedies do not eliminate the symptoms or the warning, turn off the power and contact your sales representative or local Canon dealer. Be ready to provide a detailed description of the symptoms and message.

 

Symptom

 

Cause and remedy

 

 

 

The power cord has been disconnected. Turn off the power to

The power lamp does not light

 

the retinal camera and reconnect the power cord.

when the power is turned on.

 

 

 

The power cord has been cut. Contact your sales

 

 

 

representative or local Canon dealer.

 

 

 

5 short beeps sound and the

 

A system error has occurred. Contact your sales representative

operation panel lamps blink when

 

 

or local Canon dealer.

the power is turned on.

 

 

 

 

 

 

The observation monitor is dim

 

Adjust the brightness of the observation monitor. See the Rics

and hard to see.

 

NM 2 manual.

 

 

 

The observation monitor does not

 

Turn on the digital camera.

turn on.

 

 

 

 

 

 

An error appears on the

 

A malfunction has been detected in the digital camera. Turn off

observation monitor.

 

the power to the retinal camera and then turn it on again.

Err 06

 

Sensor cleaning is not possible.

Err 70

 

Photography is not possible due to an error.

 

 

 

The working distance dots are

 

The observation light is too bright. Reduce the observation light

hard to see.

 

 

brightness.

 

 

 

The split lines are hard to see.

 

 

 

 

 

 

The working distance dots are not

 

The retinal camera and the eye are not aligned. Align them

visible.

 

correctly.

 

 

 

 

 

 

 

The retinal camera and the eye are not aligned. Align them

Flare appears in the image.

 

correctly. If flare still appears, refer to items 1 through 4 below

 

 

 

and try taking the photo again.

1)

White flare appears around

 

The retinal camera and the patient’s eye are separated. Move

 

the edges of the image.

 

the objective lens closer to the patient’s eye.

2)Orange flare appears around The retinal camera and the patient’s eye are too close. Move

 

the edges of the image.

the objective lens away from the patient’s eye.

3)

White flare appears on the

The retinal camera is too close to the left side of the patient’s

 

left side of the image.

eye. Move the objective lens to the right.

4)

White flare appears on the

The retinal camera is too close to the right side of the patient’s

 

right side of the image.

eye. Move the objective lens to the left.

 

 

 

 

 

The patient's pupil is not dilated enough. Dilate the patient’s

 

 

pupil. If the pupil is still too small, enable the small pupil

 

 

photography function.

 

 

 

 

 

Eyebrows or eyelashes are blocking the pupil. Instruct the

The split lines are not visible.

patient to open his/her eye wider.

 

The patient's eye is cloudy. Focusing cannot be performed with

 

 

 

 

the split lines.

 

 

 

 

 

The diopter compensation knob is set to either negative or

 

 

positive. Set it to the 0 position.

 

 

 

 

 

The split line lamps are not lit.

40