Appendix
EMC (Electromagnetic Compatibility)
The
If this equipment causes harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
•Reorient or relocate the receiving device.
•Increase the separation between the equipment.
•Connect this device into an outlet on a circuit different from that to which the other devices are connected.
If the problem cannot be solved with the above measures, stop using this equipment and consult your sales representative or local Canon dealer.
Precautions on EMC
1.Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the manual.
2.Portable and mobile RF communications equipment can affect medical electrical equipment.
3.Information regarding the cable affecting EMC is as follows.
To maintain the optimum EMC performance, use only the designated cables.
Name | Type | Length | Remarks |
AC Power Cord | 3.0 m | Supplied | |
|
|
|
|
USB Cable | Shielded | Max. 3.0 m | Not supplied |
|
|
|
|
HDMI Cable | Shielded | Max. 2.7 m | Not supplied |
|
|
|
|
4.The use of accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by Canon sales representative or local Canon dealer as replacement parts for internal components, may result in increased emissions or decreased immunity of the
5.The