UserManual:1528L
SW200095RevA,Page7of39
Guidance and manufacturer’s declaration – electromagnetic immunity
The 1528L is intended for use in the electromagnetic environment specified below. The customer or the user of the
1528L should assure that it is used in such an environment.
Immunity test IEC60601
test level Compliance level Electromagnetic environment –
guidance
Electrostatic discharge
(ESD)
IEC61000-4-2
±6 kV Contact
±8 kV air
±6 kV Contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material the relative humidity
should be at least 30 %
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Main power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC6100-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Main power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on the power
supply input lines
IEC61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Main power quality should be that of a
typical commercial or hospital
environment. If the user of the 1528L
requires continued operation during
power mains interruptions, it is
recommended that the 1528L be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
Magnetic field
IEC61000-4-8
3 A/m 3 A/m, 30 A/m Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE UT is the ac. mains voltage prior to application of the test level.