Guidance and manufacturer’s declaration – electromagnetic immunity

The 1528L is intended for use in the electromagnetic environment specified below. The customer or the user of the 1528L should assure that it is used in such an environment.

Immunity test

IEC60601

Compliance level

Electromagnetic environment –

test level

guidance

 

 

 

 

 

 

Electrostatic discharge

±6 kV Contact

±6 kV Contact

Floors should be wood, concrete or

(ESD)

±8 kV air

±8 kV air

ceramic tile. If floors are covered with

IEC61000-4-2

 

 

synthetic material the relative humidity

 

 

 

should be at least 30 %

 

 

 

 

Electrical fast

±2 kV for power supply

±2 kV for power supply

Main power quality should be that of a

transient/burst

lines

lines

typical commercial or hospital

IEC61000-4-4

±1 kV for input/output

±1 kV for input/output

environment.

 

lines

lines

 

 

 

 

 

Surge

±1 kV differential mode

±1 kV differential mode

Main power quality should be that of a

IEC6100-4-5

±2 kV common mode

±2 kV common mode

typical commercial or hospital

 

 

 

environment.

 

 

 

 

Voltage dips, short

<5 % UT

<5 % UT

Main power quality should be that of a

interruptions and voltage

(>95 % dip in UT)

(>95 % dip in UT)

typical commercial or hospital

variations on the power

for 0,5 cycle

for 0,5 cycle

environment. If the user of the 1528L

supply input lines

40 % UT

40 % UT

requires continued operation during

IEC61000-4-11

power mains interruptions, it is

 

(60 % dip in UT)

(60 % dip in UT)

recommended that the 1528L be

 

for 5 cycles

for 5 cycles

powered from an uninterruptible power

 

70 % UT

70 % UT

supply or a battery.

 

 

 

(30 % dip in UT)

(30 % dip in UT)

 

 

for 25 cycles

for 25 cycles

 

 

<5 % UT

<5 % UT

 

 

(>95 % dip in UT)

(>95 % dip in UT)

 

 

for 5 sec

for 5 sec

 

 

 

 

 

Power frequency

3 A/m

3 A/m, 30 A/m

Power frequency magnetic fields

(50/60 Hz)

 

 

should be at levels characteristic of a

Magnetic field

 

 

typical location in a typical commercial

IEC61000-4-8

 

 

or hospital environment.

 

 

 

 

NOTE UT is the ac. mains voltage prior to application of the test level.