Recommended separation distances between

Portable and mobile RF communications equipment and the 1528L

The 1528L is intended for use in which radiated RF disturbances are controlled. The customer or the user of the 1528L can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the 1528L as recommended below, according to the maximum output power of the communications equipment.

 

Separation distance according to frequency of transmitter

Rated maximum output power of

 

M

 

 

 

 

transmitter

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2,5 GHz

W

__

__

__

 

d = 1,2 √P

d = 1,2 √P

d = 2,3 √P

 

 

 

 

0,01

0,12

0,12

0,23

 

 

 

 

0.1

0,38

0,38

0.73

 

 

 

 

1

1,2

1,2

2,3

 

 

 

 

10

3,8

3,8

7,3

 

 

 

 

100

12

12

23

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

This device complies with all applicable electromagnetic emission and immunity standards for medical device equipment. This device is designed to not cause harmful interference, and to accept any interference received, including interference that may cause undesired operation. The performance of this device is limited to the emission and immunity standards that have been applied. Other devices which are not designed to withstand emission levels as specified in the medical device standards may be susceptible to interference from this device. Subjecting the device to conditions beyond the rated performance capabilities may result in emissions in excess of the standard. If it is determined that this device produces electromagnetic or other interference it must be disconnected from power until the cause of the problem has been determined and resolved. If it is determined that this device is functioning improperly due to electromagnetic and other interference it must be disconnected from power until the cause of the problem has been determined and resolved.

User Manual: 1528L

SW200095 Rev A, Page 9 of 39

Page 9
Image 9