Rosemount 708
Figure 5-6. Manual Alert Setup screen
If a device is installed and shows an unexpected alert, verify the installation conditions.
POWER MODULE REPLACEMENT
Expected power module life is ten years at reference conditions.(1)
When the power module needs to be replaced, remove the power module cover and the power module (SmartPower Solutions™, model number 701PGN green power module) then replace the power module and the cover. Ensure that the power module cover is fully tightened to prevent moisture ingress. The lip of the polymer power module cover should be in contact with the surface of the polymer enclosure to ensure a proper seal. Do not over tighten.
Handling Considerations
The green power module contains one “D” size primary lithium/thionyl chloride battery. Each battery contains approximately 5.0 grams of lithium. Under normal conditions, the battery materials are
Use caution when handling the power module, it may be damaged if dropped from heights in excess of 6.1 meters (20 feet).
Battery hazards remain when cells are discharged.
Environmental Considerations
As with any battery, local environmental rules and regulations should be consulted for proper management of spent batteries. If no specific requirements exist, recycling through a qualified recycler is encouraged. Consult the materials safety data sheet for battery specific information.
(1)Reference conditions are 21 °C (70 °F), transmit rate of once per minute, and routing data for three additional network devices.