Reference Manual
Rosemount 708
Section 3 | Mounting |
SAFETY MESSAGES
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page
Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the operations. Information that potentially raises safety issues is indicated by a warning symbol (). Please refer to the following safety messages before performing an operation preceded by this symbol.
Warnings
Failure to follow these installation guidelines could result in death or serious injury:
Only qualified personnel should perform the installation
Explosions could result in death or serious injury.
Before connecting a Field Communicator in an explosive atmosphere, make sure that the instruments are installed in accordance with intrinsically safe field wiring practices
Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations certifications
This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation.
This device must be installed to ensure a minimum antenna separation distance of 20 cm (8 in.) from all persons.
MOUNTING | 1. | Locate the 708 on a section of piping as close as possible to the |
|
| equipment to be monitored. Align the waveguide of the transmitter as |
|
| shown in Figure |
| 2. | The mounting location should be free of foreign matter and corrosion |
|
| to ensure good contact between the piping and the waveguide. |
| 3. | Tighten each clamp to 10.2 |
|
| band material to prevent unwanted acoustic noise. |
| 4. | If commissioning the device, install the Green Power Module (see |
|
| Figure |
| 5. | Ensure that the Power Module cover is fully tightened to prevent |
|
| moisture ingress. The lip of the polymer Power Module cover should |
be in contact with the surface of the polymer enclosure to ensure a proper seal. Do not over tighten.