Reference Manual

00809-0100-4708, Rev AD September 2012

Rosemount 708

Section 3

Mounting

SAFETY MESSAGES

Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1

Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the operations. Information that potentially raises safety issues is indicated by a warning symbol (). Please refer to the following safety messages before performing an operation preceded by this symbol.

Warnings

Failure to follow these installation guidelines could result in death or serious injury:

Only qualified personnel should perform the installation

Explosions could result in death or serious injury.

Before connecting a Field Communicator in an explosive atmosphere, make sure that the instruments are installed in accordance with intrinsically safe field wiring practices

Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations certifications

This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation.

This device must be installed to ensure a minimum antenna separation distance of 20 cm (8 in.) from all persons.

MOUNTING

1.

Locate the 708 on a section of piping as close as possible to the

 

 

equipment to be monitored. Align the waveguide of the transmitter as

 

 

shown in Figure 3-1and Figure 3-2.

 

2.

The mounting location should be free of foreign matter and corrosion

 

 

to ensure good contact between the piping and the waveguide.

 

3.

Tighten each clamp to 10.2 N-m (90 in-lb). Trim the excess clamp

 

 

band material to prevent unwanted acoustic noise.

 

4.

If commissioning the device, install the Green Power Module (see

 

 

Figure 3-3 on page 3-2).

 

5.

Ensure that the Power Module cover is fully tightened to prevent

 

 

moisture ingress. The lip of the polymer Power Module cover should

be in contact with the surface of the polymer enclosure to ensure a proper seal. Do not over tighten.

3-1

Page 17
Image 17
Emerson 708 manual Section, Mounting