5.2 I/O Module Controls 5.2.1 SIO Module – Configuration of relay outputs

Configuration of Relay Outputs:

Press the 8 -key or the -key in the line "Configuration of Relay Outputs..." of the menu "SIO - Configuration Parameters" to change to the corresponding menu:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

-- Relay Outputs --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Outputnumber:

 

 

 

 

1

 

 

 

 

 

 

 

 

Invert signal:

 

 

 

 

 

 

Disabled

 

 

 

 

 

 

 

 

Choose source module...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Choose signal...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Signal comes from:

 

 

 

 

Control Module

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Signal name:

 

 

 

 

Function control

 

 

 

 

 

 

 

 

Actual state:

 

 

 

 

 

 

 

 

 

ON

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

 

 

 

 

Back...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

F3

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

There are three relays on any SIO board. The relay contact status is NO (normally opened) by factory setting. With a jumper you also have the option to determine the relay contact status as NC (normally closed). Informations about the whole specification of the SIO board you can find in another manual.

In the menu "Relay Outputs" you have to determine the configuration of the three relay outputs of the SIO!

Configuration setup:

1)Select the output number:

·Press the 8 -key or the -key in the line "Outputnumber" and adjust the output number one, two or three with the -key or the -key.

·Confirm the number selected with the 8 -key.

2)Choose the source module:

·Change with the -key to the line "Choose source module...".

·Change to the submenu "Analyzer Modules" with the 8 -key or the -key. (illustration see next page)

All analyzer modules joint to the platform or to the MLT analyzer and the control module are available.

·Go to the line you want with the -key or the -key. If there are more than eight source modules available, you have to press the F5 key to go to another menu page.

·Select the tag of the module with the 8 -key or the -key:

The display will go back to the menu "Relay Outputs" automatically. The tag of the module selected will appear in the line "Signal comes from".

5 - 90

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98