Instruction Manual | Clarity II T56 |
May 2012 | |
|
|
3.4.4Sensor wiring to signal boards
Plug the
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
3.4.5Sensor
The sensor cable is
1.Remove the wrenching nut from the strain relief fitting.
2.Insert the plug through the hole in the bottom of the enclosure nearest the sensor socket. Seat the fitting in the hole.
3.Slide the wrenching nut over the cable plug and screw it onto the fitting.
4.Loosen the cable nut so the cable slides easily.
5.Insert the plug into the appropriate receptacle. To remove the plug, squeeze the release clip and pull straight out.
6.Adjust the cable slack in the enclosure and tighten the cable nut. Be sure to allow sufficient slack to avoid placing stress on the cable and connections.
7.Plug the cable into the back of the sensor. The sensor is rated to IP65 when properly connected to the cable. To prevent possible water damage to the connector contacts, be sure the cable receptacle and the connector on the back of the sensor are dry when connecting or disconnecting the cable.
8.Place the sensor in either the measuring chamber or the calibration cup. The sensor must be in a dark place when power is first applied to the analyzer.
NOTE:
If “S1 Warning” appears, check sensor cable connection and confirm sample water flow at debubbler drain outlet.
Wiring | 13 |