Section 4.0 Display and Operation

Instruction Manual

Clarity II T56

PN-51-T56

May 2012

 

 

Section 4.0 Display and Operation

4.1User Interface

The T56 has a large display which shows two live measurement readouts in large digits and up to six additional process variables or diag- nostic parameters concurrently. The display is back-lit and the format can be customized to meet user requirements. The ENTER/MENU key allows access to Calibration, Hold (of current outputs), Programming, Display, Data and HART functions. In addition, a dedicated INFO key is available to provide access to useful diagnostic and instrument information regarding installed sensor(s) and any problem- atic conditions. The display flashes a red banner to indicate a Fault condition and a yellow banner for a Warning condition. Help screens are displayed for fault and warning

conditions to guide the user in troubleshooting. During calibration and programming, key presses guide the user step-by-step through procedures. An alpha-numeric keypad similar to a cell phone keypad is available to allow the user to enter data during programming and calibration or lengthy tags to describe process points, sensors, or instrumentation.

4.2Instrument Keypad

There are three Function keys, four Navigation keys and an alpha-numeric keypad on the instrument keypad.

Function keys

The ENTER/MENU key is used to access menus for programming and calibrating the instru- ment as well as retrieving stored data. Eight top-level menu items appear when pressing the ENTER/MENU key from the main display of live readings:

Calibrate: calibrate attached sensors and analog outputs.

Program: Program outputs, relays, measurement, temperature, and security codes.

Hold: Suspend current outputs.

Display Setup: Program graphic trend display, brightness, main display format, tags, language, and warnings.

Data storage and retrieval: Enable data and event storage, download data, and view events.

HART or Profibus: Program HART and Profibus communication parameters.

Time and Date: Set and view real-time clock settings.

Reset: Reset all instrument settings, calibration settings or current outputs to factory defaults.

Calibrate

Program

Hold

Display setup

Data storage and retrieval

HART

Time and Date

Reset