Clarity II T56

PN-51-T56

May 2012

 

 

Section 2.0 – Installation

2.1Unpacking and Inspection

The Clarity II Turbidimeter is a complete system for the determination of turbidity in drinking water. The system consists of the analyzer, sensor(s), cable(s), and flow chamber/debubbler(s). Consult the table to verify that you have received the parts for the option you ordered.

Item

Model/part number

Single Input Turbidity Analyzer with HART

56-03-27-38-HT

Dual Input Turbidity Analyzer with HART

56-03-27-37-HT

Sensor-EPA standard

8-0108-0002-EPA

Sensor-ISO standard

8-0108-0003-ISO

Cable-3 ft (0.9 m)

24138-00

Cable-20 ft (6.1 m)

24097-00

Cable-50 ft (15.2 m)

24098-00

Calibration cup

24101-00

Molded chamber/debubbler

24170-00

2.2Installation

2.2.1General Information

1.Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperatures.

2.Install the analyzer in an area where vibration and electromagnetic and radio frequency interference are minimized or absent.

3.Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the analyzer.

4.The analyzer is suitable for panel, pipe, or surface mounting. Refer to the table below.

Type of Mounting

Figure

 

 

Panel

2-1

 

 

Wall and Pipe

2-2

 

 

2.1Unpacking and Inspection WARNING

RISK OF ELECTRICAL SHOCK

Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.

Installation