Instruction Manual | Clarity II T56 |
May 2012 | |
|
|
3.2Preparing Conduit Openings
There are six conduit openings in all configurations of T56. (Note that four of the openings will be fitted with plugs upon shipment.)
Conduit openings accept
NOTE:
Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the conduit before attaching the fitting to the analyzer.
3.3Preparing Sensor Cable
The T56 is intended for use with all Rosemount Analytical sensors. Refer to the sensor installation instructions for details on preparing sensor cables.
3.4Power, Output, and Sensor Connections
3.4.1Power wiring
Two Power Supplies are offered for the T56:
a.24VDC (20 – 30V) Power Supply
b.85 – 265 VAC Switching Power Supply
AC mains leads and 24VDC leads are wired to the Power Supply board which is mounted vertically on the left side of the main enclosure cavity. Each lead location is clearly marked on the Power Supply board. Wire the power leads to the Power Supply board using the lead markings on the board.
The grounding plate is connected to the earth terminal of the
3.4.2Current output wiring
All instruments are shipped with four
3.4.3Alarm relay wiring
Four alarm relays are supplied with the switching power supply (85 to 265VAC,
12 | Wiring |