Instruction Manual | Clarity II T56 |
May 2012 | |
|
|
Section 8.0 Maintenance
8.1Overview
The 56 Analyzer used in the Clarity II turbidimeter needs little routine maintenance. Clean the analyzer case and front panel by wiping it with a clean soft cloth dampened with water ONLY. Do not use solvents, like alcohol, that might cause a buildup of static charge.
A few of the components of the analyzer are replaceable. See Tables
WARNING
Explosion Hazard. Do not disconnect equipment when a flammable or combustible atmosphere is present.
8.2Sensor
8.2.1Cleaning the sensor
Clean the sensor by rinsing it with water followed by wiping with a soft tissue. If water is inade- quate, wash with a mild detergent solution followed by thorough rinsing with water. Do not scratch the lamp or photodiode windows.
If mineral scale is present, use a dilute acid solution applied with a cotton swab to clean away the deposit. Rinse thoroughly with water.
Do not use abrasive cleaners or solvents.
8.2.2Replacing the lamp/LED board
The
To replace the lamp/LED board:
1. Turn off power to the analyzer.
WARNING
Explosion Hazard. Do not disconnect equipment when a flammable or combustible atmosphere is present.
2. Remove the sensor from the measuring chamber and disconnect the cable.
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
Maintenance | 27 |