SPS 4001B
Instruction Manual
ELECTRICAL | All wiring must conform to local and national codes. Use the following |
CONNECTIONS | procedure to connect an SPS 4001B to an Oxymitter 4000. |
|
|
|
|
|
|
|
|
| Disconnect and lock out power before connecting the unit to the power supply. |
|
|
|
|
| NOTE |
| Ensure the Oxymitter 4000 is set up to handshake with the sequencer by |
| configuring the logic I/O to mode 8. Refer to the Oxymitter 4000 Oxygen |
| Transmitter Instruction Manual for more information. |
|
|
|
|
| NOTE |
| To maintain proper earth grounding, ensure a positive connection exists |
| between the ground lug and earth. |
|
|
1.Loosen the SPS 4001B cover screws and open the cover.
2.Route the line voltage cable into the enclosure through the rear cable grip. Ensure adequate wire length to connect the line voltage leads per Figure
3.Connect the incoming 90 to 250 VAC ±10%, 50/60 Hz line voltage leads to the terminal strip and ground lug as indicated in Figure
4.Route the signal wiring cable through the front cable grip. Ensure adequate wire length to connect all signal wires per Figure
5.Connect the 5V (5 mA maximum) logic I/O leads from the Oxymitter 4000 to the terminal strip as indicated in Figure
6.To set up the SPS 4001B to initiate a calibration from a remote location, connect the 5 VDC input leads to the terminal strip as shown in Figure
7.Relay output connections are available on the unit to signal when the Oxymitter 4000 is in calibration or when calibration failed. Relay outputs can be connected to either indicator lights or a computer interface. The relay contacts are capable of handling a 5 to 30 VDC maximum power source. The cabling requirement is 1000 ft (303 m) maximum. Connect the relay output wires to the terminal strip as shown in Figure
8.Once all connections are made, close the enclosure cover and tighten the cover screws.