The SPS 4001B is fully enclosed in a NEMA cabinet suited for wall-mounting. The enclosure provides added protection against corrosive atmospheres and incidental impact damage.

Rosemount Analytical has offered multiprobe autocalibration sequencer systems for many years. These autocalibration systems are most cost effective for boilers and other combustion processes that utilize many probes. Users with only one probe per combustion process can take advantage of Rosemount Analytical's autocalibration capability with the SPS 4001B.

SPECIFICATIONS

Specifications

SPS 4001B Assembly

6A00175G01

Part Number

 

 

 

Mounting

Wall mount

Materials of Construction:

 

Enclosure

Fiberglass

Manifold

Aluminum with 1/4 in. NPT ports

Pneumatic tubing - Low

Buna N with overbraid

Pressure

 

Pneumatic tubing - High

Impolene

Pressure

 

Assembly hardware

Galvanized and stainless steel

Humidity range

5 - 95% relative humidity, non condensing

Ambient temperature range

-40° to 149°F (-40° to 65°C) with optional heater

 

-25° to 149°F (-5° to 65°C)

Enclosure protection

NEMA 4 (IP66)

Hazardous area option(1)

Z-purge (by owner)

Electrical entries

1/2 in. NPT (0.875 in. with gland)

Input power

90 to 250 VAC ±10%, 50/60 Hz

Power consumption

5VA maximum

External electrical noise

EN 61326-1 class A

Handshake signal to/from

5V (5mA maximum)

Oxymitter 4000

 

(self-powered)

 

Cal initiate input from control

Internally 5V powered

room:

 

In-Cal Relay Contact(4)(5)

5 to 30 VDC, 100 mA, Form A (SPST)

Cal Failed Relay Contact(4)(5)

5 to 30 VDC, 100 mA, Form A (SPST)

Cabling distance between SPS

Maximum 1000 ft (303 m)

4001B and Oxymitter 4000

 

 

 

Piping distance between SPS 4001B and Oxymitter 4000 (2) (3)

Maximum 300 ft (91 m)

Approximate shipping weight

10 lbs (4.5 kg)

Pollution degree

2

Overvoltage Category

II

Certification

General Purpose

C

US

APPROVED

 

 

NOTES

 

 

(1)Hazardous area option requires the addition of a Z-purging system by the owner.

(2)Reference air is recommended with 9 ft (2.74 m) and 12 ft (3.66 m) long probes. Reference air is also recommended when ambient air may not contain the normal 20.95% O2, such as when the probe is mounted into a positive pressure duct with leaks or where there is a process unit nearby with leaks.

(3)Customer to provide piping from remote SPS 4001B to probe.

(4)External current limiting device must be provided by customer.

(5)If contacts are used to drive relay coils, a reverse blocking (flyback) diode must be installed on the input to the relay coil. Relay and diode to be provided by customer.