Emerson instruction manual SPS 4001B Troubleshooting Flowchart Sheet 2

Models: SPS 4001B

1 58
Download 58 pages 28.96 Kb
Page 33
Image 33

Instruction Manual

IM-106-340AC-1, Rev. 1.0 April 2005

SPS 4001B

Figure 5-1. SPS 4001B Troubleshooting Flowchart (Sheet 2 of 2)

SYMPTOM — NO CALIBRATION GAS FLOW (CONTINUED)

CONTINUED FROM SHEET 1 OF 2

CHECK +5 VDC AND +24 VDC. SEE NOTE 1.

+5 VDC

NO

AND +24 VDC

 

 

±10%?

 

NOTE 1: +5 VDC IS ON TP2; TP3 IS RETURN. +24 VDC IS ON TP1; TP3 IS RETURN.

NOTE 2: UNPLUG J5. CONNECT SOLENOID PINS 1 AND 3

TO +24 VDC ON TP1. TP3 IS RETURN.

REPEAT FOR SOLENOID 2 CONNECTED TO J6.

CIRCUIT BOARD IS

FAULTY. REPLACE

CIRCUIT BOARD.

YES

ADDITIONAL INFORMATION

CHECK SOLENOID 1 AND

SOLENOID 2. (SEE NOTE 2 FOR

INSTRUCTIONS).

SOLENOID

YES

CIRCUIT BOARD IS

1 AND 2

 

 

FAULTY. REPLACE

 

 

OK?

 

 

CIRCUIT BOARD.

 

NO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

REPLACE THE SOLENOID

 

 

 

WITH NO FLOW.

 

 

 

 

 

 

 

 

IF FLOW STARTS THEN STOPS AFTER SEVERAL SECONDS, THE PRESSURE SWITCH MAY BE BAD. TO TEST PRESSURE SWITCH, UNPLUG FROM J3 AND INSTALL JUMPER ACROSS J3 TERMINALS. THEN, RESTART CALIBRATION. IF THIS CORRECTS THE PROBLEM, THE PRESSURE SWITCH IS BAD; REPLACE IT.

37740009

5-5

Page 33
Image 33
Emerson instruction manual SPS 4001B Troubleshooting Flowchart Sheet 2