Reference Manual

00809-0100-4808, Rev CA June 2008

Rosemount 3051N

FIGURE 2-5. Square Root Output Transition Point.

Full Scale Output

 

 

Full Scale

(mA dc)

 

Flow (%)

 

 

 

 

 

Sq. Root

Curve

Transition Point

Linear Section

Sq. Root Curve

Transition

Point

Slope=42

Slope=1

Rerange

The Range Values command sets the 4 and 20 mA points (lower and upper

 

range values). Setting the range values to the limits of expected readings

 

maximizes transmitter performance. In practice, you may reset the transmitter

 

range values as often as necessary to reflect changing process conditions.

 

You may use one of three methods to rerange the transmitter. Each method is

 

unique; examine all three closely before deciding which method to use.

Rerange with a

Communicator Only

HART Comm.

1, 2, 3, 1, 1

 

 

Reranging using only the communicator changes the values of the analog 4 and 20 mA points independently without a pressure input.

NOTE

Changing the lower or upper range point results in similar changes to the span.

To rerange using only the communicator, enter the fast-key sequence above, select 1 Keypad input, and follow the on-line instructions. Or enter the values directly from the ONLINE screen.

2-9

Page 21
Image 21
Emerson Process Management 3051N manual Rerange with a Communicator Only

3051N specifications

The Emerson Process Management 3051N is a state-of-the-art pressure transmitter that exemplifies precision and reliability in process automation. Designed for a wide range of industrial applications, the 3051N is celebrated for its advanced features, innovative technologies, and robust construction, catering to the evolving needs of process industries.

One of the standout features of the 3051N is its exceptional accuracy. With a standard accuracy rating of ±0.075% of span, it ensures precise pressure measurements, critical for maintaining optimal process conditions. This high level of accuracy makes it suitable for applications in industries such as oil and gas, chemicals, and water treatment, where even minor deviations can lead to significant operational issues.

The 3051N employs Smart Pressure Technology, integrating digital communication capabilities with traditional analog systems. This technology allows for seamless interoperability with existing process control systems, providing users with vital data for diagnostics, calibration, and configuration via HART, FOUNDATION fieldbus, or Profibus protocols. These capabilities lead to enhanced process efficiency and reduced downtime due to effective predictive maintenance.

Another notable characteristic of this transmitter is its robust design. The 3051N is built to endure harsh industrial environments, with a rugged housing that is resistant to corrosive and hazardous conditions. Its wide operating temperature range, typically from -40°C to 85°C, coupled with various diaphragm materials, ensures that it can function effectively across diverse applications.

Moreover, the Emerson 3051N incorporates advanced predictive diagnostics, enabling users to monitor the health of the transmitter continuously. This feature alerts operators to potential issues before they escalate, facilitating proactive maintenance strategies that can significantly lower operational costs.

The flexible configuration options offered by the 3051N also enhance its usability. Users can select from a variety of pressure ranges, process connection types, and electrical connections, making it easy to customize the transmitter to fit specific application needs.

In conclusion, the Emerson Process Management 3051N pressure transmitter is a versatile and highly efficient device that integrates advanced technology with robust design and exceptional accuracy. Its ability to adapt to various industrial demands while providing predictive diagnostics makes it a preferred choice for professionals seeking reliability and precision in process management.