6.6 PEAK / HOLD MENU
The
SELECT MODE
PEAK HOLD
If neither pushbutton is depressed for fifteen (15) seconds after the above dis- play appears, the unit will automatically return to the Operating Mode accom- panied by a single “beep”.
6.6.1 PEAK
The PEAK feature allows the user to display both the lowest and highest readings that were detected since the instrument was last turned on, or since the readings were last reset.
In the SELECT MODE menu, depress the “PEAK” pushbutton. Then one of the following screens will appear for fifteen (15) seconds displaying the high- est reading since instrument turn on:
| PEAK <gas> LEVEL | |
For Toxic Gases | MAX = <_> ppm |
|
| RESET | SHOW |
| PEAK | PEAK |
| PEAK O2 LEVEL |
|
For O2 | MAX = 20.9% |
|
| MIN = 20.2% |
|
| RESET | SHOW |
| PEAK | PEAK |
If neither the “RESET PEAK” or “SHOW PEAK” pushbutton is depressed, the unit will return to the Operating Mode accompanied by a single “beep”.
If the “SHOW PEAK” pushbutton is depressed, the instrument will return to the Operating Mode and display the peak values as follows on the appropriate screen:
23