6.7 ALARM SETPOINT MENU

The “ALARM SETPOINT” can be adjusted by the user in the ALARM SETPOINT menu within the following ranges:

CO

= 5 to 250 ppm

Cl2

= 0.1 to 10.0 ppm

ClO2 = 0.10 to 2.00 ppm HCN = 0.1 to 50.0 ppm H2S = 5 to 100 ppm NH3 = 5 to 100 ppm NO2 = 0.2 to 5.0 ppm O2 LO = 0.0 to 21.0% O2 HI = 21.0 to 25.0% PH3 = 0.10 to 5.00 ppm SO2 = 0.5 to 20.0 ppm

Very low oxygen alarm settings are useful when the instrument is used to verify nitrogen purging. On leaving the setpoint function or on subsequent power on, the user will be prompted to verify the safety of abnormally low settings

To enter the ALARM SETPOINT menu, depress the “SELECT” pushbutton in the menu selection screen as described in Section 6.2 above. One of the following ALARM SETPOINT menu screens will appear:

 

<gas> ALARM SETPOINT

For Toxic Gases

HI ALARM = <_> ppm

 

CHANGE

OK

 

ALARM SETPOINT

For O2

HI ALARM = 23.5%

 

LO ALARM = 19.5%

 

 

CHANGE

OK

The setpoint shown is the factory default setting. If acceptable, depress

26