CARBON MONOXIDE SPECIFICATIONS
Range:
0 to 500 ppm (for CO 500)
0 to 999 ppm (for CO 999) Response Time:
T90 < 25 seconds Operating Temperature Range:
Graphic Display Temperature Range:
(+14 to +140 oF)
Operating Humidity:
15 to 90% RH,
0to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours -
Sensor Cross Sensitivity:
Filtered against H2S, 15 ppm: < 0.5 ppm Filtered against SO2, 5 ppm ≈ 0 ppm Nitric Oxide, 35 ppm: < 3 ppm Nitrogen Dioxide, 5 ppm: ≈ - 1 ppm Chlorine, 1 ppm ≈ 0 ppm
Hydrogen, 100 ppm: < 40 ppm
Ethylene, 100 ppm: < 50 ppm
Ethanol, 200 ppm: ≈ 0 ppm
Consult Factory for additional information.
40