Energizer PORTABLE GAS DETECTION INSTRUMENT manual Disable Battery Saver MODE? Yesno

Models: PORTABLE GAS DETECTION INSTRUMENT

1 50
Download 50 pages 37.92 Kb
Page 33
Image 33

If an alarm condition occurs while in the BATTERYSAVER mode, the graphic display will come on automatically and remain on until the alarm is reset. After the alarm indication clears, the graphic display will remain on for thirty (30) seconds and then the display will go blank as described above.

If the BATTERY SAVER mode had already been enabled , the graphic display will show the following:

DISABLE BATTERY

SAVER MODE?

YESNO

Depressing the “YES” pushbutton will turn off the BATTERY SAVER mode.

6.9.2 PASSWORD PROTECTION

The PASSWORD PROTECTION menu is intended to be used by authorized per- sonnelforthepurposeoflockingoutthemenustopreventchangestotheinstrument settings.Afour digit numeric password can be set up to restrict access to all menus exceptAUTO CALIBRATION or ZERO CALIBRATION. To use this feature, de- press the “SELECT” pushbutton when the CONFIGURATION MENU screen is dis- played as described in section 6.2. The following screen will appear:

CONFIGURATION

BATTERY PASSWORD

SAVER MENU

Depress the “PASSWORD MENU” pushbutton and the following screen will appear:

ENTER PASSWORD 0000

UPSET

Use the “UP” and “SET” push-buttons to enter the correct password. (The MINI-SA comes from the factory with the default password: 1234.) Entering an incorrect password will display an INCORRECT PASSWORD message and allow the user to enter the password again. When the correct password

33

Page 33
Image 33
Energizer PORTABLE GAS DETECTION INSTRUMENT Disable Battery Saver MODE? Yesno, Password Protection, Enter Password Upset