Energizer PORTABLE GAS DETECTION INSTRUMENT manual Sensor Replacement Message, Ment Message

Models: PORTABLE GAS DETECTION INSTRUMENT

1 50
Download 50 pages 37.92 Kb
Page 22
Image 22

Note: SCOTT INSTRUMENTS supplies Teflon®-lined tubing with all chlorine and ammonia calibration kits. Sticky gases such as chlorine and ammonia require this tubing be used for calibration. Using tub- ing of another type may inhibit the automatic calibration of a MINI-SA and lead to improper readings or a false failed sensor message. All other calibration kits are supplied with Tygon (R-3603) Tubing.

SCOTT INSTRUMENTS also recommends that the tubing and cylinder regulator used with a specific toxic gas not be used with other gases.

6.5 SENSOR REPLACEMENT MESSAGE

If a successful “AUTO CALIBRATION” reveals that the gas sensor is near the end of its service life, the MINI-SA will display one of the following REPLACE-

MENT MESSAGE:

 

WARNING <gas> SENSOR

For Toxic Gases

APPROACHING FAILURE

 

REPLACE SOON!

 

ACKNOWLEDGE

 

WARNING OXYGEN SENSOR

For O2

APPROACHING FAILURE

 

REPLACE SOON!

 

ACKNOWLEDGE

The user must “ACKNOWLEDGE” this message by depressing either pushbutton.

WARNING

THIS SCREEN IS FOR INFORMATION ONLYAND IS NOT INTENDED TO BE AN EXACT DETERMINATION OF WHEN A SENSOR WILL FAIL AND REQUIRE REPLACEMENT. THE TIME REMAINING UNTIL A SENSOR FAILS AFTER A REPLACEMENT MESSAGE IS DISPLAYED IS BASED ON MANY FACTORS SUCH AS FREQUENCY OF CALIBRATION, LAST CALIBRATION, FREQUENCY OF USE AND LEVEL OF GAS TO WHICH SENSOR IS EXPOSED. A SENSOR MAY FAIL PRIOR TO DISPLAY OF THE REPLACEMENT MESSAGE.