CHLORINE SPECIFICATION

Range:

0.0to 99.9 ppm

Response Time:

T80 < 60 seconds Operating Temperature Range:

-20 to +50 oC (-4 to +122 o F)

Graphic Display Temperature Range: -10 to +60 oC

(+14 to +140 o F)

Operating Humidity:

15 to 90% RH, Non-Condensing

Recommended Storage Temperature:

0to 20 oC (32 to 68 oF)

Sensor Accuracy:

±3% of Applied Gas

Linearity:

±3% of Full Scale

Battery Life:

About 100 Hours - Non-Alarming

About 800 Hours in Battery Save Mode

Sensor Cross Sensitivity:

Carbon Monoxide at 300 ppm: 0 ppm

Hydrogen Sulfide, 15 ppm: -3 to 0 ppm

Sulphur Dioxide, 5 ppm: 0 ppm

Nitric Oxide, 35 ppm: 0 ppm

Consult Factory for additional information.