Energizer PORTABLE GAS DETECTION INSTRUMENT manual Hydrogen Cyanide Specification

Models: PORTABLE GAS DETECTION INSTRUMENT

1 50
Download 50 pages 37.92 Kb
Page 47
Image 47

HYDROGEN CYANIDE SPECIFICATION

Range:

100ppm

Response Time:

T90 < 200 seconds Operating Temperature Range:

-20oto +50 oC (-4oto +122o F)

Graphic Display Temperature Range: -10 to +60 oC

(+14 to +140 o F)

Operating Humidity:

15to 90% RH, Non-Condensing

Recommended Storage Temperature:

0to 20 oC (32 to 68 oF)

Sensor Accuracy:

±3% of Applied Gas

Linearity:

±3% of Full Scale

Battery Life:

About 100 Hours - Non-Alarming About 800 Hours in Battery Save Mode

Sensor Cross Sensitivity:

Carbon Monoxide at 300 ppm: <15 ppm

Hydrogen Sulfide, 15 ppm: 90 ppm

Sulpher Dioxide, 20 ppm: <75 ppm

Nitric Oxide, 35 ppm: <0 ppm

Nitrogen Dioxide, 5 ppm: < -10 ppm

Ethylene, 100 ppm: <25 ppm

Consult Factory for additional information.