HYDROGEN CYANIDE SPECIFICATION
Range:
100ppm
Response Time:
T90 < 200 seconds Operating Temperature Range:
Graphic Display Temperature Range:
(+14 to +140 o F)
Operating Humidity:
15to 90% RH,
Recommended Storage Temperature:
0to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours -
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: <15 ppm
Hydrogen Sulfide, 15 ppm: ≈ 90 ppm
Sulpher Dioxide, 20 ppm: <75 ppm
Nitric Oxide, 35 ppm: <0 ppm
Nitrogen Dioxide, 5 ppm: <
Ethylene, 100 ppm: <25 ppm
Consult Factory for additional information.
47