Extron electronic FPM 101, FLS 101 manual Ref/Set Button, Splash and Auto-Off Screen

Models: FPM 101 FLS 101

1 30
Download 30 pages 58.05 Kb
Page 13
Image 13
Ref/Set Button

Ref/Set Button

The Ref/Set button, Ref  , provides two functions:

Press and hold to store the power level values for a set of reference cables, (Set),

Press and release to display previously stored references, (Ref).

Test Mode: To Set a Reference

1.Refer to the sections describing connection of reference cables (see “Testing Fiber Optic Links” on page 12).

2.Set the FPM 101 to the desired measurement mode (dBm or uW) (see “dB/dBm/μW Button” in the previous section).

3.To store the currently measured level or multiple levels (in dual WAVE ID mode) as the new reference level, press and hold the Ref/Set button until HELDSET is displayed.

4.Once the new reference is set, the FPM 101 switches to insertion loss (dB) mode.

Test Mode: To Set a Reference NOTE: When setting references from dual WAVE ID sources, additional time is

required for the power meter to identify and measure both wavelengths. Allow Test Mode: To View the Reference several seconds before pressing the Ref/Set button in step 3.Splash and Auto-Off Screen

Test Mode: To View the Reference

1.Press the Ref/Set button to display the stored reference wavelength and power level for the currently selected wavelength or detected WAVE ID wavelengths. The reference indicator (A, below) lights, and the reference values are displayed briefly (several seconds) as follows:

 

 

 

 

 

 

 

 

nm

 

 

 

 

 

 

 

 

C

 

 

 

 

 

 

 

µW

 

REF

dBmdBm

 

 

 

WAVE ID

nm

 

 

 

 

 

 

 

 

 

 

 

µW

 

 

 

 

 

 

 

 

dBmdBm

Manual backgroundD

Manual backgroundE

Single and Dual WAVE ID mode

Continuous Wave (CW) mode — The reference wavelength and power level are displayed at B

Single WAVE ID mode — The reference wavelength and power level are displayed at B

Dual WAVE ID mode — The first reference wavelength and power level are displayed at B,and the second at C .

Splash and Auto-Off Screen

Splash (Version) Screen — During power up, the FPM 101 software version is shown on the top line, (D). The FPM 101 type and detector type are shown on the next line (E). The splash screen is replaced by the normal display after about

nm

µW

REF

dB dBm

F G

two seconds.

Auto-off Screen — When the unit is about to power down due to inactivity, the display shows OFF for a brief time on line E, then counts down the number of seconds until power is shut off.

WAVE ID

Hz nm

µW

dB dBm

Fiber Optic Test Meters • Operation

7

Page 13
Image 13
Extron electronic FPM 101, FLS 101 manual Ref/Set Button, Splash and Auto-Off Screen, Test Mode To Set a Reference