Mandrels for Multimode Transmit Reference Cables

When testing multimode links using an overfilled LED source, always wrap the transmit reference cable five (5) times around the proper diameter mandrel. This is specified by TIA/EIA-568-B and improves insertion loss measurement repeatability and accuracy. Do NOT use mandrels on multimode receive reference cables or singlemode reference cables.

Available mandrels for test

Diameter (D)

reference cables with 3 mm jackets

for other jacket

 

 

diameters (in mm)

Fiber Type

Part Number

 

 

 

 

62.5 μm

Contact Extron

D = 20 mm minus

 

 

jacket diameter

50 μm

Contact Extron

D = 25 mm minus

 

 

jacket diameter

Kit with both

Contact Extron

 

mandrels

 

 

 

 

 

Multimode Transmit Jumper

D

(5) Wraps

Mandrel

Loops for Singlemode Transmit Reference Cables

For TIA testing of singlemode links, the transmit reference cable must have a 1.2 inch

(30 mm) loop secured by tape, hook and loop fasteners, or other means. Do not pull the fastener too tight.

Transmit

Reference Cable

1.2inch (30 mm) Diameter Loop

Secure

Loop Here

FLS 101

Fiber Optic Test Meters • Maintenance 19

Page 25
Image 25
Extron electronic FPM 101 Mandrels for Multimode Transmit Reference Cables, Loops for Singlemode Transmit Reference Cables