6.3REPLACING FUSES
WARNING:
Be sure to replace fuses with the correct type and rating, specified in Table 4.
Always replace fuse F3 with a 2 A fast blow type as serious injury could result from the wrong fuse. Open the case and remove the PCB (refer to § 6.2)
Note: A limited number of MR810AEA units were manufactured with fuse F3 rated at 1.5 A. These units can be identified by the fuse rating information
1.Slide the cable clamp up out of the transformer mount and lift it away from the fuses.
2.The fuses can now be accessed. Refer to Figure 6.3 for the location of the fuses and Table 4 for replacement fuses.
3.Install the PCB and assemble the case (refer to § 6.8).
Fuse F1
Fuse F2
6
Fuse F3
Figure 6.3: Fuse LocationTable 4: Replacement fuse ratings and part numbers (for single fuses)Model | Supply Voltage | Fuse Type | Part Number |
230 V~ | F1: 1.5 A 250 V~, Fast Blow | 999 830 008 | |
|
| F2: 1.5 A 250 V~, Fast Blow | 999 830 008 |
|
| F3: 2.0 A 250 V~, Fast Blow | 999 830 009 |
|
|
|
|
115 V~ | F1: 2.0 A 250 V~, Fast Blow | 999 830 009 | |
|
| F2: 2.0 A 250 V~, Fast Blow | 999 830 009 |
|
| F3: 2.0 A 250 V~, Fast Blow | 999 830 009 |
|
|
|
|
100 V~ | F1: 2.0 A 250 V~, Fast Blow | 999 830 009 | |
|
| F2: 2.0 A 250 V~, Fast Blow | 999 830 009 |
|
| F3: 2.0 A 250 V~, Fast Blow | 999 830 009 |
|
|
|
|